Rad-5/5v Signal Extraction Pulse Oximeter Operator’s Manual 8-5
8
sensors & patient cables
Masimo SET patient cables
Reusable patient cables of various lengths are available. Only use appropriate Masimo
oximetry patient cables for SpO
2
measurements. Other patient cables may cause improp-
er Rad-5/5v handheld pulse oximeter performance.
CLEANING AND REUSE OF MASIMO SET PATIENT CABLES
Patient cables can be cleaned per the following procedure:
■
Remove the cable from the sensor.
■
Disconnect the cable from the monitor.
■
Wipe clean with a 70% isopropyl alcohol pad.
■
Allow the cable to dry before returning it to operation.
CAUTIONS:
■
CAREFULLY ROUTE PATIENT CABLES TO REDUCE THE POSSIBILITY OF
PATIENT ENTANGLEMENT OR STRANGULATION.
■
DO NOT SOAK OR IMMERSE PATIENT CABLES IN ANY LIQUID SOLUTION. DO
NOT STERILIZE PATIENT CABLES BY IRRADIATION, STEAM, OR ETHYLENE
OXIDE. SEE THE CLEANING INSTRUCTIONS IN THE DIRECTIONS FOR USE
FOR REUSABLE MASIMO PATIENT CABLES.
■
DO NOT REPROCESS ANY MASIMO SET PATIENT CABLES.
LNOPv
TM
ADHESIVE SENSORS
(LNOPv sensors must be used in conjunction with PC cables)
CLEANING AND REUSE OF MASIMO SENSORS
Reusable sensors can be cleaned per the following procedure:
■
Remove the sensor from the patient.
■
Disconnect the sensor from the monitor.
■
Wipe the entire sensor clean with a 70% isopropyl alcohol pad.
■
Allow the sensor to air dry before returning it to operation.
REATTACHMENT OF SINGLE USE ADHESIVE SENSORS
■
Single use sensors may be reapplied to the same patient if the emitter and detector
windows are clear and the adhesive still adheres to the skin.
■
The adhesive can be partially rejuvenated by wiping with a 70% isopropyl alcohol
wipe and allowing the sensor to thoroughly air dry prior to replacement on the patient.
NOTE: If the sensor fails to track the pulse consistently, the sensors may be incorrectly
positioned. Reposition the sensor or choose a different monitoring site.
CAUTIONS:
■
DO NOT REPROCESS ANY SINGLE USE SENSORS.
■
DO NOT SOAK OR IMMERSE THE SENSOR IN ANY LIQUID SOLUTION.DO NOT STER-
ILIZE ANY MASIMO SENSOR BY IRRADIATION, STEAM, OR ETHYLENE OXIDE.
WARNING: TO AVOID CROSS CONTAMINATION ONLY USE MASIMO SINGLE
USE SENSORS ON THE SAME PATIENT.
Weight Saturation Accuracy Pulse Rate Accuracy
Low Perfusion Accuracy
SENSOR Range No Motion Motion No Motion Motion Saturation Pulse Rate
LNOPv In 3 - 20 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
LNOPv Ne < 3 kg ± 3% ± 3% ± 3 bpm ± 5 bpm ± 3% ± 3 bpm
LNOPv Ad > 30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
LNOPv Ad-L > 30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
LNOPv Pd-L 10-50 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
sensors & patient cables
8-4 Rad-5/5v Signal Extraction Pulse Oximeter Operator’s Manual