Rad-5/5v Signal Extraction Pulse Oximeter Operator’s Manual 4-15
4
Trend Setup and Use (Rad-5 only)
INTRODUCTION
The Rad-5 can store 72 hours of SpO
2
and Pulse Rate and Perfusion Index trend data,
captured at 2 second intervals. This trend data can then be transferred to a PC for evalu-
ation.
Trend data is stored in non-volatile memory, so it is not erased when the unit is shut off or
when the batteries are replaced.
A special serial cable (see Section 10,
Accessories
) is required to connect the sensor con-
nector of the Rad-5 to the PC. Patient monitoring is not possible while trend memory is
being transferred to a PC.
A trend data download is initiated using the TrendCom utility which downloads the trend
data and saves it to a space-delimited ASCII text (.out) file.
TRENDCOM UTILITY INSTALLATION
Copy the TrendCom utility from the CD onto a PC running MS-Windows.
TRENDCOM UTILITY OPERATION
1. Disconnect patient sensor and/or cable from the Rad-5.
2. Connect the mini-D end of the PRONTO serial cable to the Rad-5 patient cable
connector (see Section 2,
Rad-5 front panel controls
) and connect the DB-9 end to
a COM port on the PC.
3. Turn the Rad-5 on
4. Start the TrendCom Utility
5. Select the appropriate COM port number, if necessary.
6. Push the RETRIEVE TREND button on the TrendCom utility.
Select the desired location and assign a filename for the trend file. Press Save.
7. The Rad-5 will display “dat out” while trend data is being transferred.
A progress bar will advance to indicate the status of the download.
Larger trend files will take longer to download. Transfer time is approximately 20
seconds per hour of trend data.
Note: During download of trend information, all normal Rad-5 functions are unavail-
able and the keypad is locked, except for the power button.
8. When trend data transfer is complete, close TrendCom and disconnect the Rad-5
from the PRONTO serial cable.
9. Turn the Rad-5 off to exit the trend download mode.
Note: USB to serial port adapters are not supported for trend transfer.
Note: Enabling trend (setting Trend to “ON”) will erase all trend information in the Rad-5.
operation
4-14 Rad-5/5v Signal Extraction Pulse Oximeter Operator’s Manual
4
ERASING TREND MEMORY
To erase (clear) the trend memory, turn the trend off and back on again. Enabling trend
(setting Trend to “ON”) will erase all trend data.
Note: Turning trend off will not erase trend memory. You may turn trending off and still
retrieve the trend data using TrendCom.
Turning the Rad-5 off or replacing the batteies will not erase the trend data.
Turn trending off before storing the unit for any length of time.
TREND DATA FORMAT
After a successful download of the trend data, a .out file will be created containing the
trend-dump information in ASCII delimited format. The format is defined in the following
table.
SAMPLE TREND OUTPUT
07/21/04 09:56:08 SpO2=000 PR=000 PI=00.00 EXC=820:OffPat,SET
07/21/04 09:56:10 SpO2=000 PR=000 PI=00.00 EXC=828:Search,OffPat,SET
07/21/04 09:56:12 SpO2=097 PR=069 PI=04.69 EXC=800:SET
07/21/04 09:56:14 SpO2=096 PR=074 PI=02.28 EXC=C00:LowSigIQ,SET
07/21/04 09:56:16 SpO2=098 PR=078 PI=03.64 EXC=800:SET
07/21/04 09:56:18 SpO2=000 PR=000 PI=00.00 EXC=800:SET
07/21/04 09:56:20 SpO2=000 PR=000 PI=00.00 EXC=820:OffPat,SET
07/21/04 09:56:22 SpO2=096 PR=078 PI=02.68 EXC=800:SET
PARAMETER SPECIFICATION
Date MM\DD\YY
Time HH:MM:SS
SpO2 001 to 100, or "---" meaning parameter not available
Pulse Rate 001 to 240, or "---" meaning parameter not available
Perfusion Index 00.00 to 20.00
Exception
Messages
The exceptions are displayed as a 3 digit, ASCII encoded, hexadecimal
value. The binary bits of the hexadecimal value are encoded as follows:
000 = Normal operation; no exceptions
001 = No Sensor
002 = Defective Sensor
004 = Low Perfusion
008 = Pulse Search
010 = Interference
020 = Sensor Off
040 = Ambient Light
080 = Unrecognized Sensor
100 = reserved
200 = reserved
400 = Low Signal IQ
800 = Masimo SET. This flag means the algorithm is running is full SET
mode. It requires a SET sensor and needs to acquire some clean
data for this flag to be set