Rad-5/5v Signal Extraction Pulse Oximeter Operator’s Manual 4-3
4
operation
DEFAULT SETTINGS
The Rad-5/5v oximeters store two types of default values: those that the device automati-
cally reverts to after a power cycle, and those that can be changed by the user and will be
remembered after a power cycle.
The following table outlines the default values that the Rad-5 and Rad-5v revert to after a
power cycle:
The following table outlines the default values that the Rad-5 reverts to after a power cycle:
*
Defaults to APOD and Normal only. High Sensitivity will default to normal.
OPTION DEFAULT SETTING
SpO
2
high alarm limit
Set to Off
SpO
2
low alarm limit
Set to 90%
Pulse rate high alarm limit Set to 140 BPM
Pulse rate low alarm limit Set to 50 BPM
Averaging Time Set to pre-power down setting
FastSat Set to pre-power down setting
Sensitivity* Set to pre-power down setting
Display brightness Set to pre-power down setting
Pulse tone volume Set to pre-power down setting
Alarm Silence Set to all audible alarms active
Alarm Volume Set to pre-power down setting
Trending on/off
Set to pre-power down setting
(Strongly recommend turning trending off
prior to turning unit off)
Sleep Study Mode Set to pre-power down setting
OPTION DEFAULT SETTING
Display brightness Set to pre-power down setting
Pulse tone volume Set to pre-power down setting
pulse rate alarm limits beyond the patient readings.
■
An alarm tone sounds.
■
The violated alarm limit and reading flash on the display.
11. Verify the sensor alarms are functional by removing the sensor from the sensor
site.
■
“SEn OFF” message appears on the display.
■
The alarm tone sounds.
■
The Visual Alarm Indicator flashes.
■
Disconnect the sensor from the patient cable or oximeter.
■
Confirm that “NO SEn” message appears on the display.
Note: “NO SEn” and “SEn OFF” will only generate an alarm if the Rad-5/5v was
actively monitoring a patient when the sensor was disconnected.
12. Rad-5 only:
Verify parameter-violation alarm silence operation.
■
Create an alarm condition by lowering the SpO
2
or pulse rate high alarm lim-
its beyond the patient readings.
■
Press the Alarm Silence button.
■
The alarm tone ceases for 120 seconds.
13. To begin patient monitoring:
■
Adjust the alarm limits (Rad-5 only).
■
Adjust the alarm volume (Rad-5 only).
■
Adjust the pulse beep volume.
15. Verify the sensor is applied correctly and that the measured data is appropriate,
see Section 4,
Successful SpO
2
Monitoring
.
16. Monitor the patient.
17. After monitoring is complete, remove the sensor from the patient and store or dis-
pose of the sensor according to governing rules. See the Directions for Use of the
sensor. If trending is enabled, turn off.
18. Press and hold the Power/Standby Button for 2 seconds to turn the oximeter off.
Note: Turn the oximeter off between patients so that it can re-calibrate in order to interpret
new physiological data and to conserve battery life.
operation
4-2 Rad-5/5v Signal Extraction Pulse Oximeter Operator’s Manual