Rad-5/5v Signal Extraction Pulse Oximeter Operator’s Manual 4-3
4
operation
DEFAULT SETTINGS
The Rad-5/5v oximeters store two types of default values: those that the device automati-
cally reverts to after a power cycle, and those that can be changed by the user and will be
remembered after a power cycle.
The following table outlines the default values that the Rad-5 and Rad-5v revert to after a
power cycle:
The following table outlines the default values that the Rad-5 reverts to after a power cycle:
*
Defaults to APOD and Normal only. High Sensitivity will default to normal.
OPTION DEFAULT SETTING
SpO
2
high alarm limit
Set to Off
SpO
2
low alarm limit
Set to 90%
Pulse rate high alarm limit Set to 140 BPM
Pulse rate low alarm limit Set to 50 BPM
Averaging Time Set to pre-power down setting
FastSat Set to pre-power down setting
Sensitivity* Set to pre-power down setting
Display brightness Set to pre-power down setting
Pulse tone volume Set to pre-power down setting
Alarm Silence Set to all audible alarms active
Alarm Volume Set to pre-power down setting
Trending on/off
Set to pre-power down setting
(Strongly recommend turning trending off
prior to turning unit off)
Sleep Study Mode Set to pre-power down setting
OPTION DEFAULT SETTING
Display brightness Set to pre-power down setting
Pulse tone volume Set to pre-power down setting
pulse rate alarm limits beyond the patient readings.
An alarm tone sounds.
The violated alarm limit and reading flash on the display.
11. Verify the sensor alarms are functional by removing the sensor from the sensor
site.
“SEn OFF” message appears on the display.
The alarm tone sounds.
The Visual Alarm Indicator flashes.
Disconnect the sensor from the patient cable or oximeter.
Confirm that “NO SEn” message appears on the display.
Note: “NO SEn” and “SEn OFF” will only generate an alarm if the Rad-5/5v was
actively monitoring a patient when the sensor was disconnected.
12. Rad-5 only:
Verify parameter-violation alarm silence operation.
Create an alarm condition by lowering the SpO
2
or pulse rate high alarm lim-
its beyond the patient readings.
Press the Alarm Silence button.
The alarm tone ceases for 120 seconds.
13. To begin patient monitoring:
Adjust the alarm limits (Rad-5 only).
Adjust the alarm volume (Rad-5 only).
Adjust the pulse beep volume.
15. Verify the sensor is applied correctly and that the measured data is appropriate,
see Section 4,
Successful SpO
2
Monitoring
.
16. Monitor the patient.
17. After monitoring is complete, remove the sensor from the patient and store or dis-
pose of the sensor according to governing rules. See the Directions for Use of the
sensor. If trending is enabled, turn off.
18. Press and hold the Power/Standby Button for 2 seconds to turn the oximeter off.
Note: Turn the oximeter off between patients so that it can re-calibrate in order to interpret
new physiological data and to conserve battery life.