Rad-5/5v Signal Extraction Pulse Oximeter Operator’s Manual 4-13
4
operation
SETUP MENU LEVEL 4 - LED BRIGHTNESS AND FACTORY DEFAULTS
Push the Mode/Enter button again to enter menu level 4.
Pressing a fifth time returns the Rad-5 to patient monitoring in the Saturation/Pulse
Rate Mode. Additionally, the Rad-5 will automatically return to patient monitoring display
from any menu level/setting after 10 seconds with no key presses.
SETTNG
4X
LED Display Brightness (4 levels)
Note: All LED indicators are illumi-
nated while adjusting this setting.
Use
Up
or
Down
Arrow Keys to
adjust parameter to desired setting.
Restore Factory Defaults
Yes/No
Save User Identified Default Settings
(Password: next button, up arrow,
down arrow, next button)
SETUP MENU LEVEL 3 - TREND SETTINGS
Push the Mode/Enter button again to enter menu level 3.
To enable trending of patient data, the trend feature must be enabled (set to ON), and the
current date and time must be set. See section 4,
Trend setup and use
.
The current date and time can only be set if the Trend is set to “ON”. The date and time
menu selections are not available if Trend is set to “OFF”.
A valid date must be entered. If an invalid date is entered (i.e. February 31), the trend
will not turn on and “tnd off will be displayed.
Note: Press and hold for rapid scrolling:
SAT alarm keys up / down will scroll numbers.
PR alarm keys up / down will scroll numbers.
Note: The date and time must be set before trending will be enabled. The Rad-5 will
automatically ‘time out’ of the setup menu after 10 seconds with no key presses. If the
Rad-5 should time-out fo the Trend Settings menu, the trend will not
be enabled.
Note: Enabling trend (setting Trend to “ON”) will erase all trend information in the Rad-
5. Turn trending off prior to turning unit off and storing,
SETTNG
3X
Trend ON / OFF
Use
Up
key to turn trend ON.
Use
Down
key to turn trend OFF
Set Year
Use
Up
or
Down
Arrow Keys to
adjust parameter to desired setting.
Set Month
Set Day
Set Hour
Set Minute
operation
4-12 Rad-5/5v Signal Extraction Pulse Oximeter Operator’s Manual