RAD-5 FRONT PANEL CONTROL OPERATION
Setup menu (Rad-5 only)
This section gives an overview of the Rad-5 menu selections available. These options do
not apply to the Rad-5v. To navigate through the menus, use the
Mode/Enter
,
Next
,
Up
and
Down
keys located on the front panel of the oximeter, below the LED display. The follow-
ing sub-sections describe each menu item in more detail. The oximeter has options that
allow user configuration to suit specific needs.
MENU NAVIGATION
The Rad-5 set-up and configuration options are accessed through the menu system. The
Mode/Enter
key is used to enter the menu system and to move through the different menu
levels. Within each level of the system, the
Next
key is used to move from one option to
the next. The
Up
and
Down
arrow keys are used to select values within each option. The
parameter is set/selected when either the
Mode/Enter
or
Next
keys are pressed.
BUTTON FUNCTION
Power on/off. Press to turn Rad-5 on. Press-and-hold for 2 seconds to turn
Rad-5 off.
Enters the Rad-5 setup/menu system. See Section 4,
Setup menu
.
No function during normal patient monitoring.
Alarm Silence. Pressing this button one time will temporarily silence a satura-
tion or pulse rate high/low limit violation for 120 seconds. A second press will
permanently silence all audible alarms.
Pressing this button will acknowledge and permanently silence a ‘sensor-off
and ‘no-sensor’ audible alarm. It will also permanently silence a low battery
audible alarm if the Rad-5 is not monitoring a patient.
If a low battery alarm occurs duing patient monitoring, pressing the Alarm
Silence button will silence the audible alarm for 120 seconds.
During normal patient monitoring the Up and Down Arrow keys control the
Pulse Tone volume. At the lowest setting, the pulse tone is muted. A low-pitch
tone indicates the highest or lowest setting has been reached.
In the setup/menu system, the Up and Down Arrow keys select among the
options for each setting.
operation
4-10 Rad-5/5v Signal Extraction Pulse Oximeter Operator’s Manual
4
Rad-5/5v Signal Extraction Pulse Oximeter Operator’s Manual 4-11
4
operation
SETUP MENU LEVEL 1 – ALARM LIMITS AND ALARM VOLUME.
SETUP MENU LEVEL 2 – AVERAGING AND SENSITIVITY
Push the Mode/Enter button again to enter menu level 2.
1
Defaults to APOD and Normal only. High sensitivity will default to normal.
2
FastSat is automatically enabled in 2 and 4 second averaging.
SETTNG
2X
Averaging.The signal averaging time
of this device can be set to:
2*, 4*, 8, 10, 12, 14 or 16 seconds
Use
Up
or
Down
Arrow Keys to
adjust parameter to desired setting.
Note: The parameter is set/select-
ed when the Mode Enter and Next
are pressed.
Sensitivity
1
.
Hi = Maximum
Nor = Normal
APO = APOD
FastSat
2
On, Off
SETTNG
Alarm Volume
Use
Up
or
Down
Arrow Keys to
adjust parameter to desired setting.
Note: The parameter is set/selected
when the Mode Enter and Next are
pressed.
SpO
2
High Alarm Limit
SpO
2
Low Alarm Limit
Pulse Rate High Alarm Limit
Pulse Rate Low Alarm Limit