Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual 5-3
5
MESSAGE POSSIBLE CAUSE(S) RECOMMENDATION
REPLACE
ADHESIVE
When a single patient use sensor is used,
the adhesive portion of the sensor is
non-functional, or the life of the adhesive
portion of the sensor has expired. (Applies
to Rainbow ReSposable CO-Oximeter
Sensor Systems, or ReSposable Pulse
Oximeter Sensor Systems only.)
Replace the adhesive portion of the
sensor.
REPLACE
REUSABLE
The reusable sensor (i.e., the
durable portion that includes the
emitter and detector) is non-
functional, or the life of the reusable
sensor has expired. (Applies to
Rainbow ReSposable CO-Oximeter
Sensor Systems, or ReSposable
Pulse Oximeter Sensor Systems
only).
Replace the reusable sensor.
REPLACE
CABLE
The patient cable is non-functional,
or the life of the cable has expired.
Replace the patient cable.
INTERFERENCE
High intensity light such as pulsating
strobe lights, excessive ambient
light sources such as surgical lights
or direct sunlight, or other monitor
displays.
Place a Masimo Optical Light Shield
over the sensor.
Incorrect monitor line frequency
setting (Hz).
Access the Traditional User Interface
described in Section 4. Select Confi g
and enter the password. Adjust the
Line Frequency to the correct Hz
setting.
CHECK
SENSOR
Sensor is not connected fi rmly into
patient cable, or the sensor is not
connected fi rmly to the instrument.
Reconnect sensor fi rmly into patient
cable, or to the instrument.
SPO2 ONLY
MODE
"SpO2 Only Mode" message occurs
during an unsuccessful sensor
calibration/pulse search routine, or
during monitoring.
Review the sensor's Directions
for Use instructions, Section 4,
Successful Monitoring, and Section 8,
Selecting a Masimo SET Sensor.
LOW
BATTERY
Battery charge is low.
Charge battery by placing the
Radical-7 Handheld into the Docking
Station and powering the instrument
with AC line power. Replace battery if
necessary.
LOW
PERFUSION
Signal too small.
Move sensor to better perfused site.
Refer to Section 4, Low Perfusion.
Alarms and Messages