4-24 Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual
4
Operation
Output (continued)
MENU ITEMS DESCRIPTION
ANALOG1 OR
ANALOG 2
(CONTINUED)
0V OUTPUT
A 0 Volt calibration signal is mapped to the analog output. Use this
signal for calibration of recording instruments. (0 Volts represents a
saturation of 0% and a pulse rate of 0 bpm).
1V OUTPUT
A 1 Volt calibration signal is mapped to the analog output. Use this
signal for calibration of recording instruments. (1 Volt represents a
saturation of 100% and a pulse rate of 250 bpm).
NURSE CALL
ALARMS
The nurse call output will be activated based on alarm events.
LOW SIGNAL IQ
The nurse call output will be activated based on Low Signal IQ events.
ALARM & SIGNAL IQ
The nurse call output will be activated based on alarm and Low Signal
IQ events.
POLARITY
NORMAL
Standard polarity. See Section 7, Analog output / nurse call
specifi cations.
INVERT
This setting reverses the Normally Open and Normally Closed
contacts. See Section 7, Analog output / nurse call specifi cations.
CAUTION: TO AVOID EXCESSIVE BATTERY DISCHARGING, DO NOT CONNECT ANY
EQUIPMENT TO THE SERIAL PORT ON THE BACK PANEL UNLESS THE
RADICAL-7 IS CONNECTED TO THE AC POWER FROM THE WALL OUTLET.