9-2 Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual
9
Battery Operation and Maintenance
The Radical-7 Pulse CO-Oximeter Handheld includes a 1.5 Amp-Hour Nickel Metal Hydride
battery. The Radical-7 Pulse CO-Oximeter Docking Station may include the optional 6.5 Amp-
Hour Nickel Metal Hydride battery.
Before using the Radical-7 Pulse CO-Oximeter as a Handheld or transport monitor, the
Handheld battery and the optional Docking Station battery need to be fully charged.
To charge the battery(s), attach the Handheld instrument to the Docking Station. Ensure that AC
power is attached to the Docking Station. Verify that the battery(s) is charging; the battery charging
LED indicators on the Docking Station flash prior to charging and remain illuminated while the
battery(s) is charging. A continuously flashing battery charging LED indicates that the internal
battery temperature exceeds recommended operating conditions for proper battery charging. Proper
battery charging will proceed when the temperature returns to recommended operating conditions.
The Handheld battery requires approximately 2 to 3 hours for charging. The optional Docking
Station battery requires approximately 6 hours for charging.
When the battery charging LED indicators turn off, additional trickle charging may occur to complete
charging. Although battery charging can occur while the Handheld is docked and powered on, most
efficient charge times are achieved with the Handheld instrument turned off.
CAUTIONS:
ALL BATTERIES LOSE CAPACITY WITH AGE, THUS THE AMOUNT OF RUN TIME
LEFT AT LOW BATTERY WILL VARY DEPENDING UPON THE AGE OF THE BATTERY.
AT LOW BATTERY CONNECT THE RADICAL-7 PULSE CO-OXIMETER TO AC
POWER TO PREVENT LOSS OF POWER.
During battery operation of the Radical-7 Pulse CO-Oximeter, please note that the following
operating conditions affect the estimated run-time of the included batteries:
ILLUMINATION OF THE BACKLIT LCD SCREEN. TO CONSERVE BATTERY POWER,
KEEP THE BACKLIT LCD SCREEN AT MINIMUM ILLUMINATION.
VOLUME OF THE ALARM TONES. TO CONSERVE BATTERY POWER, KEEP THE
FREQUENCY OF THE AUDIBLE ALARMS TO A MINIMUM AND AT MINIMUM VOLUME.
THE SATSHARE FEATURE. TO CONSERVE BATTERY POWER, ALWAYS KEEP THE
INSTRUMENT ON AC LINE POWER.
Memory effects of the battery pack may shorten run-time. When battery run-time is significantly reduced,
it is advisable to completely discharge and fully recharge the battery pack. To properly discharge the
battery pack, use the Battery Discharge function as described in Section 4, under Service.
CAUTION:
IF THE RADICAL-7 PULSE CO-OXIMETER HANDHELD HAS NOT BEEN USED OR
CHARGED WITHIN SEVEN (7) DAYS OR MORE, THEN RECHARGE THE BATTERY
PRIOR TO USE.
IT IS RECOMMENDED THAT THE RADICAL-7 PULSE CO-OXIMETER HANDHELD IS
DOCKED TO THE DOCKING STATION ATTACHED TO AN AC POWER SOURCE WHEN
IT IS NOT IN USE TO ENSURE THAT THE BATTERY REMAINS FULLY CHARGED.