Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual 2-3
2
System Description
HANDHELD FRONT PANEL (CONTINUED)
1
HANDHELD
RELEASE
BUTTON
Press down the Handheld Release Button and pull the Handheld
instrument off the Docking Station.
2
97
SpO
2
MEASUREMENT
DISPLAY
The functional arterial hemoglobin oxygen saturation is
displayed in units of percentage SpO
2
. The upper and lower
SpO
2
alarm limits are also displayed next to the SpO
2
measurement. When a sensor is not connected to a patient and
during pulse search, the display will show dashed lines and
the message "Sensor Off" will appear at the top of the display
screen. When the measured value is outside of the alarm limits,
the SpO
2
measurement display flashes and an alarm will sound.
The oxygen saturation is calculated and the display is updated
at a frequency of once per second.
3
Masimo Rainbow SET
MASIMO SET
OR MASIMO
RAINBOW SET
The Masimo SET or Masimo rainbow SET label is shown on the
Radical-7 display when either SET processing is active.
NOTE: Patient information, not SET activity, will be displayed when using
a RadNet
®
compatible instrument connected to a monitoring system.
4
99
85
SATURATION
ALARM LIMITS
DISPLAY
The Saturation Alarm Limits Display shows the upper and lower
saturation alarm limits. When an alarm limit is exceeded, the
SpO
2
value and the violated limit flashes.
5
ALARM STATUS
INDICATOR
The alarm status indicator (a bell) can be shown with or without
a slash. It flashes when an alarm condition is present. When the
alarm is silenced using the Alarm Silence Button, an alarm status
indicator with a slash and a timer is shown to indicate that the alarm
is temporarily silenced. When the alarm is silenced through the All
Mute menu selection (which is permanent until power is cycled or
deselected using the menu), or when the Interface Alarm SpO
2
/BPM
is set to "No", an alarm status indicator with a slash is shown to
indicate the alarm has been silenced.
6
1.6
SpMet*
MEASUREMENT
DISPLAY
The measurement of methemoglobin concentration levels is
displayed in units of percentage SpMet. The upper and lower
SpMet alarm limits are also displayed next to the SpMet
measurement. When a sensor is not connected to a patient and
during pulse search, the display will show dashed lines and
the message "Sensor Off" will appear at the top of the display
screen. The methemoglobin is calculated and the display is
updated at a frequency of once per second.
7
3.0
0.3
SpMet ALARM
LIMITS DISPLAY
The SpMet Alarm Limits Display shows the upper and lower alarm
limits. When the measured value is outside of the alarm limits, the
SpMet measurement display flashes and an alarm will sound.
8
Sensor Off
SYSTEM
MESSAGE AREA
T
he system messages generated by the instrument are displayed
in the System Message Area. See Section 5, System Messages.
9
Philips VueLink
SERIAL OUTPUT
MODE
The Serial Output Mode displays the selected output interface
with the Docking Station is connected to an external instrument
via the serial port. During SatShare, the Output Mode will display
the type of SatShare cable.
10
20
SpCO*
MEASUREMENT
DISPLAY
The measurement of carbon monoxide concentration levels is
displayed in units of percentage SpCO. When a sensor is not
connected to a patient and during pulse search, the display will show
dashed lines and the message Sensor Off will appear at the top of the
display screen. The carboxyhemoglobin is calculated and the display
is updated at a frequency of once per second.
* For instruments that include SpMet, SpCO and SpHb parameters: the SpMet, SpCO, SpHb and SpOC parameters/measurements
will be dimly lit in the display screen if a patient cable is not attached or a non-Rainbow sensor is being used.
NOTE: The Pulse CO-Oximeter display view will be different without the optional SpCO, SpMet and
SpHb
parameters/measurements.