Radical Signal Extraction Pulse Oximeter Operator’s Manual 23
4
operation
successful SpO
2
monitoring
The following general points will aid in ensuring oximetry monitoring success.
Place the sensor on a site that has sufficient perfusion and provides proper
alignment of the LEDs and photodetector.
Place the sensor on a site that has unrestricted blood flow.
Do not constrict the monitoring site when securing a sensor with tape.
Do not select a site near potential electrical interference (electrosurgical unit, for
example).
Read the sensor Directions for Use for proper sensor application.
NUMERIC DISPLAY - SpO
2
Stability of the SpO
2
readings may be a good indicator of signal validity. Although stability
is a relative term, experience will provide a good feeling for changes that are artifactual or
physiological and the speed, timing, and behavior of each.The stability of the readings over
time is affected by the averaging mode being used.The longer the averaging time, the more
stable the readings tend to become. This is due to a dampened response as the signal is
averaged over a longer period of time than during shorter averaging times. However, longer
averaging times delay the response of the oximeter and reduce the measured variations of
SpO
2
and PR.
Inaccurate measurements may be caused by:
Significant levels of dysfunctional hemoglobin (e.g., carboxyhemoglobin or
methemoglobin).
Intravascular dyes such as indocyanine green or methylene blue.
Venous pulsations at the frequency of the patients arterial pulse.
Very low hemoglobin levels.
NUMERIC DISPLAY - PULSE RATE
The Pulse Rate displayed on the Radical may differ slightly from the heart rate displayed on
ECG monitors due to differences in averaging times. There may also be a discrepancy
between cardiac electrical activity and peripheral arterial pulsation. Significant differences
may indicate a problem with the signal quality due to physiological changes in the patient or
one of the instruments or application of the sensor or patient cable.The pulsations from intra-
aortic balloon support can be additive to the pulse rate displayed on the pulse oximeter .