Strany 1 - OPERATOR’S MANUAL

Signal Extraction Pulse OximeterOPERATOR’S MANUAL

Strany 2

ix Radical Signal Extraction Pulse Oximeter Operator’s Manual

Strany 3

about this manualThis manual explains how to set up and use the Radical Pulse Oximeter. Important safety information relating to general use of the Ra

Strany 4

overview2 Radical Signal Extraction Pulse Oximeter Operator’s Manual1warnings, cautions and notesPlease read and follow any warnings, cautions and not

Strany 5

product descriptionThe Radical Pulse Oximeter is a noninvasive, arterial oxygen saturation and pulse ratemonitor. Radical can be used as either a Hand

Strany 6

pulse oximetryGENERAL DESCRIPTIONPulse oximetry is a continuous and non-invasive method of measuring the level of arterialoxygen saturation in blood.

Strany 7

FUNCTIONAL VS. FRACTIONAL SATURATIONThe Radical measures and displays functional saturation: the amount of oxygenatedhemoglobin expressed as a percent

Strany 8

introductionRadical provides the functionality of three pulse oximeters in one: Radical is a fully featured Handheld pulse oximeter Radical is a fully

Strany 9

radical handheldThe Handheld Radical provides most of the functionality of the pulse oximeter. All userinput and displays are controlled by this part

Strany 10

system description8 Radical Signal Extraction Pulse Oximeter Operator’s Manual2FastSatMasimo SETPress down the Handheld Release Button and pull the Ha

Strany 11

19Radical Signal Extraction Pulse Oximeter Operator’s Manual 92system descriptionPulseWaveformDisplayTime and DateIndicatorBacklight/ContrastButtonAla

Strany 12

The Radical Operating Instructions are intended to provide the necessary information for properoperation of all models of the Radical pulse oximetry s

Strany 13

system description10 Radical Signal Extraction Pulse Oximeter Operator’s Manual231251 0000 536MEDICAL ELECTRICAL EQUIPMENTUL 2601-1/CAN/CSA C22.2 No.

Strany 14 - Patient Cable

Radical Signal Extraction Pulse Oximeter Operator’s Manual 11radical standaloneWhen the Radical Handheld is placed into the Docking Station, the Radic

Strany 15 - Masimo SET DST

system description12 Radical Signal Extraction Pulse Oximeter Operator’s Manual2STANDALONE FRONT PANELThe following figure and corresponding text revi

Strany 16

Radical Signal Extraction Pulse Oximeter Operator’s Manual 132system descriptionSTANDALONE BACK PANELP2 P3Manufactured by:Masimo CorporationIrvine, CA

Strany 17 - 06:10:23pm 03/27/01

14 Radical Signal Extraction Pulse Oximeter Operator’s Manual2SYMBOLSThe following symbols are found on the back of the Radical Docking Station, they

Strany 18

Radical Signal Extraction Pulse Oximeter Operator’s Manual 153introductionBefore the Radical Pulse Oximeter can be used in a clinical setting, it need

Strany 19

16 Radical Signal Extraction Pulse Oximeter Operator’s Manual3 IF THERE IS ANY DOUBT ABOUT THE INTEGRITY OF THE PROTECTIVE EARTHCONDUCTOR ARRANGEMENT,

Strany 20 - HANDHELD BACK PANEL

Radical Signal Extraction Pulse Oximeter Operator’s Manual 173monitor setupThe Radical Pulse Oximeter stores two types of default values: those that R

Strany 21

18 Radical Signal Extraction Pulse Oximeter Operator’s Manual3ALARM VOLUME Set to pre-power down settingPULSE BEEP VOLUME Set to pre-power down settin

Strany 22 - Standalone device

Radical Signal Extraction Pulse Oximeter Operator’s Manual 1934. Verify that Radical recognizes the correct cable. The name of the SatShare cable will

Strany 23 - STANDALONE BACK PANEL

SAFETY INFORMATION,WARNINGS, CAUTIONS AND NOTESThe Radical Signal Extraction Pulse Oximeter is designed to minimize the possibility of hazards fromerr

Strany 24 - Docking Station

set-up20 Radical Signal Extraction Pulse Oximeter Operator’s Manual3hp vuelink setup1. Select the HP VueLink selection from the Output menu on the Rad

Strany 25

Radical Signal Extraction Pulse Oximeter Operator’s Manual 214introductionTo operate the Masimo SET Radical Pulse Oximeter effectively, the device mus

Strany 26

22 Radical Signal Extraction Pulse Oximeter Operator’s Manual411. Verify that the patient alarms are functional by setting the high and low SpO2andpul

Strany 27

Radical Signal Extraction Pulse Oximeter Operator’s Manual 234operationsuccessful SpO2monitoringThe following general points will aid in ensuring oxim

Strany 28

operation24 Radical Signal Extraction Pulse Oximeter Operator’s Manual4SIGNAL IQThe Radical display provides a visual, indicator of the plethysmogram

Strany 29

Radical Signal Extraction Pulse Oximeter Operator’s Manual 254operationLOW PERFUSIONThe Radical displays an indicator (i.e., “Low Perfusion” message)

Strany 30

26 Radical Signal Extraction Pulse Oximeter Operator’s Manual4touch key control buttons and iconsThe touch key control buttons are the four dark grey

Strany 31

Radical Signal Extraction Pulse Oximeter Operator’s Manual 274navigating the main menuWhen the main menu is accessed, the plethysmograph and Signal IQ

Strany 32

28 Radical Signal Extraction Pulse Oximeter Operator’s Manual4EDITING A PARAMETEROnce a parameter has been selected for editing, a new set of menu sel

Strany 33

Radical Signal Extraction Pulse Oximeter Operator’s Manual 294operationCLOCKABOUTOUTPUTSERVICETime (hour/minute)Time display formatDate (day/month/yea

Strany 34

SAFETY INFORMATION,WARNINGS, CAUTIONS AND NOTES (CONTINUED) The pulse oximeter can be used during defibrillation, but the readings may be inaccurate f

Strany 35 - Anesthesiology

30 Radical Signal Extraction Pulse Oximeter Operator’s Manual4operationalarmsCheck alarm limits each time the pulse oximeter is used to ensure that th

Strany 36

Radical Signal Extraction Pulse Oximeter Operator’s Manual 314operationThis menu allows the user to set the alarm silence period. Analarm is silenced

Strany 37

32 Radical Signal Extraction Pulse Oximeter Operator’s Manual4operationMENU ITEMS DESCRIPTIONThree views are available: Pleth and Signal IQ, Numbers a

Strany 38

Radical Signal Extraction Pulse Oximeter Operator’s Manual 334operationgeneralAVERAGING TIME The signal averaging time of this device can be set to: 2

Strany 39

34 Radical Signal Extraction Pulse Oximeter Operator’s Manual4operationaboutThis displays the copyright and software versions of the Handheld and Dock

Strany 40

Radical Signal Extraction Pulse Oximeter Operator’s Manual 354operationCAUTION: TO AVOID EXCESSIVE BATTERY DISCHARGING, DO NOT CONNECT ANYEQUIPMENT TO

Strany 41

36 Radical Signal Extraction Pulse Oximeter Operator’s Manual4operationtrend displayOnce the Trend Display touch key icon is selected the trend data i

Strany 42 - + SIGNAL IQ

Radical Signal Extraction Pulse Oximeter Operator’s Manual 374operationThe top line on the trend display shows the time scale of the trend graph, foll

Strany 43

38 Radical Signal Extraction Pulse Oximeter Operator’s Manual4operationNAVIGATING THE TREND DISPLAYIn the Trend Display view there are a total of 10

Strany 44

Radical Signal Extraction Pulse Oximeter Operator’s Manual 394operationTREND SETUPThis menu allows the user to set the default trend settings and to c

Strany 45

table of contentsSECTION 1 / OVERVIEWAbout this Manual ...

Strany 46

40 Radical Signal Extraction Pulse Oximeter Operator’s Manual4TREND PERIODThe Trend period setting determines how often a set of SpO2and pulse rate da

Strany 47

Radical Signal Extraction Pulse Oximeter Operator’s Manual 414 The user can access the menu system. If the SatShare cable is connected to the Radical

Strany 48

42 Radical Signal Extraction Pulse Oximeter Operator’s Manual4home mode operationThe Radical can be placed into the Home Mode to protect unqualified u

Strany 49

Radical Signal Extraction Pulse Oximeter Operator’s Manual 435introductionThe following sections outline system alarms, messages displayed on the LCD

Strany 50

44 Radical Signal Extraction Pulse Oximeter Operator’s Manual5alarms/messagesalarm operationFor averaging modes of less than 14 seconds, the Radical i

Strany 51

Radical Signal Extraction Pulse Oximeter Operator’s Manual 455alarms/messagessystem messagesThe following chart alphabetically lists all system messag

Strany 52

46 Radical Signal Extraction Pulse Oximeter Operator’s Manual5alarms/messages* The SERVICE REQUIRED message fills the entire display.NO SENSOR Sensor

Strany 53

Radical Signal Extraction Pulse Oximeter Operator’s Manual 476troubleshootingtroubleshootingThe following chart describes what to do if the Radical sy

Strany 54

48 Radical Signal Extraction Pulse Oximeter Operator’s Manual7specificationsradical specificationsPERFORMANCEmeasurement rangeSpO2: 1-100%Pulse Rate:

Strany 55

Radical Signal Extraction Pulse Oximeter Operator’s Manual 497specificationsENVIRONMENTALOperating Temperature: 41°F to 104°F (5°C to 40°C)Storage Tem

Strany 56

SECTION 4 / OPERATIONIntroduction ...

Strany 57

50 Radical Signal Extraction Pulse Oximeter Operator’s Manual7specificationscomplianceEMC Compliance: EN60601-1-2, Class BEquipment Classification: IE

Strany 58 - ELECTRICAL

Radical Signal Extraction Pulse Oximeter Operator’s Manual 517specificationsPARAMETER SETTINGBAUD RATE 9600 Baud bi-directionalNUMBER OF BITS PER CHAR

Strany 59 - PHYSICAL CHARACTERISTICS

52 Radical Signal Extraction Pulse Oximeter Operator’s Manual7specificationsanalog output / nurse call specificationsThe Analog Out and Nurse Call are

Strany 60

Radical Signal Extraction Pulse Oximeter Operator’s Manual 537specificationsANALOG OUTPUTRadical can interface with various analog recording devices a

Strany 61

54 Radical Signal Extraction Pulse Oximeter Operator’s Manual8sensor/patient cablesintroductionThis section covers the use and cleaning of Masimo LNOP

Strany 62 - Station

Radical Signal Extraction Pulse Oximeter Operator’s Manual 558sensor/patient cablesCLEANING AND REUSE OF MASIMO LNOP SENSORSReusable sensors can be cl

Strany 63 - 980mV in 3.9mV steps

56 Radical Signal Extraction Pulse Oximeter Operator’s Manual8sensors/patient cablesmasimo SET patient cablesReusable patient cables of various length

Strany 64

Radical Signal Extraction Pulse Oximeter Operator’s Manual 579service/maintenanceintroductionThis chapter covers how to test the operation of the Radi

Strany 65

58 Radical Signal Extraction Pulse Oximeter Operator’s Manual9service/maintenancebattery operation and maintenanceThe Radical Handheld includes a 1.5

Strany 66

Radical Signal Extraction Pulse Oximeter Operator’s Manual 599service/maintenanceCAUTION: IF THE RADICAL HAS NOT BEEN CHARGED FOR 1 MONTH OR MORE, THE

Strany 67

Home Use...33Interface Alarms ...

Strany 68

60 Radical Signal Extraction Pulse Oximeter Operator’s Manual9service/maintenanceWARNING: THE DOCKING STATION BATTERY SHOULD BE INSTALLED AND/ORREMOVE

Strany 69

Radical Signal Extraction Pulse Oximeter Operator’s Manual 619service/maintenanceREPLACING THE FUSESShould a power problem blow one or both of the fu

Strany 70

62 Radical Signal Extraction Pulse Oximeter Operator’s Manual9service/maintenancePower-On Self-Test:1. Connect the monitor to AC power and verify that

Strany 71

Radical Signal Extraction Pulse Oximeter Operator’s Manual 639service/maintenanceTesting with Masimo SET Tester (Optional):1. Turn the Radical off and

Strany 72

64 Radical Signal Extraction Pulse Oximeter Operator’s Manual9service/maintenanceAnalog Output Test1. Disconnect all patient cables and sensors from t

Strany 73

Radical Signal Extraction Pulse Oximeter Operator’s Manual 659service/maintenanceservice and repairREPAIR POLICYMasimo or an authorized Service Depart

Strany 74

service/maintenance66 Radical Signal Extraction Pulse Oximeter Operator’s Manual9warrantyMasimo warrants to the initial purchaser that each new pulse

Strany 75

Radical Signal Extraction Pulse Oximeter Operator’s Manual 6710accessoriesACCESSORIES PART NUMBERRADICAL HANDHELD 1309RADICAL DOCKING STATION RDS-1 13

Strany 76

accessories68 Radical Signal Extraction Pulse Oximeter Operator’s Manual10ACCESSORIES PART NUMBERSATSHARE CABLE-DO01 1324SATSHARE CABLE-DO02 1326SATSH

Strany 77 - Extraction pulse oximeter:

Backlight / Contrast Operation.......40SatShare Operation

Strany 78 - 11601E-0801

Radical Signal Extraction Pulse Oximeter Operator’s Manual viiitable of contentsSECTION 9 / SERVICE AND MAINTENANCEIntroduction ......