4. Verify that Radical recognizes the correct cable. The name of the SatShare cable will
be displayed on the LCD screen when the SatShare mode is functional.
5. Set the multiparameter monitor’s high and low saturation and pulse rate alarm limits
as appropriate.
6. Set the multiparameter monitor's averaging time to the lowest setting (i.e. fastest
response). Radical's ideal waveform obviates the need for additional averaging by
the monitor. If the multiparameter monitor's averaging time is not changed, the time
to display physiological changes in saturation on the monitor will be increased with
SatShare. However, the delay can be minimized by reducing the multiparameter
monitor's averaging time.
7. While in the SatShare mode, if there are any significant discrepancies between
the readings from the Radical and those on the monitor displaying the values
obtained from SatShare, the values reported by the Radical are to be considered
the correct values.
8. To use the Radical with SatShare while it is not connected to AC power, set the
Power Save parameter in the General menu to "No", refer to Section 4, Operation.
Please note that if the Radical is used in this mode, the length of time the Radical
can operate on battery power will be significantly diminished.
9. Set the SatShare Numbers and the Interface Alarms parameters in the General
menu according to Customer preference.
10. If displaying the simulated waveform is not desirable, it is recommended to turn off
the pleth waveform display of the multiparameter patient monitor.
CAUTIONS:
■
WHILE OPERATING IN SATSHARE MODE, THE AUDIBLE ALARMS MAY BE
DISABLED ON THE RADICAL. USE THE ALARM SETTINGS OF THE
MULTIPARAMETER MONITOR FOR AUDIBLE AND VISUAL ALARM INDICATION.
■
TO AVOID EXCESSIVE BATTERY DISCHARGING, DO NOT CONNECT ANY
EQUIPMENT TO THE SATSHARE CONNECTOR UNLESS THE RADICAL PULSE
OXIMETER IS CONNECTED TO THE AC MAIN POWER SUPPLY.
■
TO ENSURE THAT THE SAFETY LEVEL DURING DEFIBRILLATION IS
MAINTAINED, MAKE SURE THAT THE SATSHARE CABLE CONNECTOR IS
PROPERLY SECURED AT THE DOCKING STATION.
set-up
WARNING: EXTERNAL DEVICE CONNECTIONS TO THE SATSHARE PORT
MUST BE IEC-60101-1-1 COMPLIANT.