Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 8-3
8
LNOP
®
REUSABLE SENSORS
LNOP
reusable sensors must be used in conjunction with Red PC cables.
SENSOR
Weight
Range
Saturation Accuracy Pulse Rate Accuracy Low Perfusion Accuracy
No Motion Motion No Motion Motion Saturation Pulse Rate
LNOP DCI > 30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
LNOP DCIP 10 - 50 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
LNOP YI
(Adults/Pediatrics)
> 1 kg
± 2%
± 3% ± 3 bpm ± 5 bpm N/A N/A
LNOP YI
(Neonates)
± 3%
LNOP TC-I > 30 kg ± 3.5% N/A ± 3 bpm N/A ± 3.5% ± 3 bpm
LNOP DC-195 > 30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
LNOP TF-I > 30 kg ± 2% N/A ± 3 bpm N/A ± 2% ± 3 bpm
NOTE: The LNOP TF-I and TC-I sensors were not validated under motion conditions.
LNOPv
TM
ADHESIVE SENSORS
LNOPv adhesive sensors must be used in conjunction with Red PC cables
SENSOR
Weight
Range
Saturation Accuracy Pulse Rate Accuracy Low Perfusion Accuracy
No Motion Motion No Motion Motion Saturation Pulse Rate
LNOPv In 3 - 20 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
LNOPv Ne < 3 kg ± 3% ± 3% ± 3 bpm ± 5 bpm ± 3% ± 3 bpm
LNOPv Ad > 30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
LNOP
®
SPECIALTY SENSORS
LNOP specialty sensors must be used in conjunction with Red PC cables
SENSOR
Weight
Range
Saturation Accuracy Pulse Rate Accuracy Low Perfusion
Accuracy
No Motion Motion No Motion Motion Saturation Pulse Rate
LNOP Newborn Infant
(thumb or great toe)
Infant (finger or toe)
3 - 30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
10 - 30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 3% ± 3 bpm
LNOP Newborn
Neonatal
< 3 kg ± 3% ± 3% ± 3 bpm ± 5 bpm ± 3% ± 3 bpm
LNOP Trauma > 30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
LNOP Blue 2.5 - 30 kg
60 - 80% ± 4% N/A ± 3 bpm N/A ± 3% ± 3 bpm
70 - 100% ± 3.3% N/A ± 3 bpm N/A ± 3% ± 3 bpm
80 -100% ± 3% N/A ± 3 bpm N/A ± 3% ± 3 bpm
sensors & patient cables