Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 4-9
4
operation
Normal patient monitoring
FRONT PANEL CONTROL OPERATION
SETUP MENU
This section gives an overview of the Rad-8 menu selections available. To access the
menu levels and navigate through the menu selections, use the front panel buttons, Enter
Button and Up/Down Buttons as indicated in the following sections. Sub-sections describe
each menu item in more detail. The Rad-
8 has options that allow user configuration to
accommodate specific needs.
MENU NAVIGATION
The Rad-8 set-up and configuration options are accessed through the menu system. Four
levels of menus are available to the user. Once a menu level is accessed, a front panel
button (Level 1 only) or the Enter Button (Level
2 and 3) is used to move from one option
to the next allowing repeated cycling through the options. The Up and Down Buttons are
used to adjust values within each option. The parameter
/measurement value is set when
the Enter Button is pressed.
When accessing the Rad-8 menu, each selection will be
communicated visually on the LED (front of device).
NOTE: The Rad-8 will automatically ‘time out’ of any setup menu after 10 seconds if no
button is pressed.
SETUP MENU LEVEL 1
ALARM LIMITS
To access alarm limits for param
eters/measurements, press the Alarm Limits Button to access
the Alarm Limits menu.
BUTTONS SETTINGS INSTRUCTIONS
Use the Alarm
Limits Button to
access the alarm
limits options and
move between
options.
Press once
%SpO
2
LO
Use the Up Button to move
between settings and the Enter
Button to accept the setting
and move to the next menu
screen
OR
let the device time out for 10
seconds to exit without saving
the new setting and to return to
the home display screen.
.
Press 2x
%SpO
2
HI
Press 3x
Pulse rate (BPM) LO
Press 4x
Pulse rate (BPM) HI