3-2 Radical-7 Color Display Signal Extraction Pulse Oximeter Operator’s Manual Radical-7 Color Display Signal Extraction Pulse Oximeter Operator’s Manual 3-3
INITIAL BATTERY CHARGING
Before use, the Radical-7 Pulse CO-Oximeter Handheld battery and the optional
Docking Station battery needs to be fully charged.
To charge the batteries:
1. Attach the Handheld unit to the Docking Station.
2. Plug in the AC power cord to power entry module. Make sure it is securely plugged
in.
3. Plug the AC power cord into an AC power source.
4. Verify that the batteries are charging.
The battery charging LED indicators on the Docking Station flash prior to charging and
remain illuminated while the batteries are charging.
Refer to Section 9, Battery Operation and Maintenance, for proper battery charging.
INITIAL INSTALLATION
Place the Docking Station on a stable hard flat surface near the patient. Always place the
Radical-7 Pulse CO-Oximeter unit on a dry surface. Maintain a minimum of 3 cm (1 inch)
free space around the Radical-7 Pulse CO-Oximeter Standalone unit. Make sure that the
Radical-7 loudspeaker is not covered to avoid a muffled alarm sound.
The Radical-7 Pulse CO-Oximeter Handheld, Docking Station or Standalone should not
be operated outside the following environmental conditions:
OPERATING ENVIRONMENTAL CONDITIONS
TEMPERATURE
+5°C to +40°C, +41°F to +104°F
HUMIDITY
5% to 95%, non-condensing
OPERATING ALTITUDE
1060 mbar to 500 mbar pressure
-1000 ft to 18,000 ft (-304 m to 5,486 m)
Configure the unit for your regional power line frequency (50 or 60 hz) if needed. Default is
60 hz (standard for the United States). See Section 4, Operation, Config.
CAUTION: THE UNIT MUST BE CONFIGURED TO MATCH YOUR LOCAL POWER
LINE FREQUENCY TO ALLOW FOR THE CANCELLATION OF NOISE
INTRODUCED BY FLUORESCENT LIGHTS AND OTHER SOURCES.
setup
3-2
3
3
setup
Monitor Setup
The Radical-7 Pulse CO-Oximeter maintains three types of default values which will
automatically revert to after a power cycle:
■ Factory – these options are restored to factory values
■ Custom – these settings can be changed by the user and retained through the
power cycle.
■ Adult / Neo – these settings can be selected to revert to factory or hospital-defi ned
values (for Adult or Neonatal) after a power cycle.
The following outlines the default values the Radical-7 Pulse CO-Oximeter reverts to
after a power cycle.
FACTORY DEFAULT SETTINGS
OPTION DEFAULT SETTING
LCD SCREEN ILLUMINATION
AC Power
Battery Power
Set to maximum, level 4
Set to minimum, level 1
SENSITIVITY
Set to normal mode
CUSTOM (USER) DEFINED SETTINGS
This mode is indicated by “Mode Custom” on the Alarms menu.
The following table outlines the settings that may be changed by the user and the Radical-7
Pulse CO-Oximeter will remember after a power cycle.
OPTION DEFAULT SETTING
DISPLAY VIEW
Set to Pleth + SigIQ
AVERAGING TIME
Set to 8
FAST S AT
Set to No
HOME USE
Set to No
INTERFACE ALARMS
Set to Yes
SATSHARE NUMBERS
Set to Yes
POWER SAVE
Set to No
DATE FORMAT
Set to mm/dd/yy
TIME FORMAT
Set to 12 hr.
LANGUAGE
Set to English
ANALOG OUTPUT
Set to Analog 1: SpO
2
0-100%
Set to Analog 2: Pulse Rate
SERIAL OUTPUT PORT MODE
Set to Binary
PULSE BEEP VOLUME
Set to Level 4
TREND DISPLAY PARAMETERS
Set to %SpO
2